Sample¶
The diamond anvil cell (the high-pressure sample environment) and the DAC positioning stage / micro-diffractometer.
Enclosure: 13-ID-D.
Note
The DAC binds the catalog PressureCell Family (HP-1). It is one Asset presenting the Regulator Role for its membrane pressure (the PACE5000 setpoint + readback); its double-sided laser heating (HEAT-1) and its in-situ pressure / temperature metrology are capabilities of the same cell. The DAC positioning stage binds the catalog Goniometer (the i03 Smargon precedent; the single Omega plus the Galil-versus-XPS controller are per-Asset settings, SAMPLE-1).
| Name | Family | PV | Key specs | Replaceable | Status |
|---|---|---|---|---|---|
PressureCell |
PressureCell |
13IDD_PACE5000:PC1: |
the diamond anvil cell: membrane gas pressure set and read on the PACE5000 (PC1:Setpoint / Pressure_RBV / Control / Vent), presenting the Regulator Role; double-sided IPG YLR laser heating (13IDD:Laser1 / Laser2, power on 13IDD:US_LaserPower / DS_LaserPower) is the cell's heating capability (HEAT-1); the in-situ temperature (spectroradiometry, 13IDD:us_las_temp / ds_las_temp) and pressure (ruby / Raman / Brillouin) are its metrology Sensor capability. catalog Family, graduated on the rule-of-three across 13-ID and PETRA III P02 (PRESSURE-1) | new confirm |
|
SampleStage |
Goniometer |
confirm: the DAC positioning stage / micro-diffractometer (Galil m1-m4 X / Z / Y / Omega, plus the Newport XPS-16 PSO trajectory stage); single Omega and the Galil-vs-XPS controller are settings (SAMPLE-1) |
the DAC sample-positioning stage with the Omega rotation; binds Goniometer (the i03 Smargon precedent); single Omega and the Galil-vs-XPS controller are settings (SAMPLE-1) | new confirm |
|
SampleTable |
Table |
confirm: the DAC lift table the cell rides on; binds Table (SAMPLE-1) |
the DAC lift table supporting the cell (SAMPLE-1) | new confirm |
|
MetrologySpectrometer |
Camera |
13IDDLF1: |
the LightField (PIMAX / PIXIS) spectrometer reading the thermal-emission and ruby / Raman spectra for the cell's in-situ temperature and pressure metrology; reuses Camera as the metrology detector (HP-1) | new confirm |