Sample¶
The TomoWISE sample stage. Design-phase; values are TDR design targets.
The sample stage is two experiment stations in the experiment hutch, sharing one Detector gantry: a microtomography station (~45 m) and a nanotomography station (~49 m). They are modelled as two sample-stage groups in the descriptor, each presenting the specimen to the beam in a different way. Models named "(target)" are the TDR design selections, carried unbound until procurement confirms them.
Unlike the Detector, the sample stage is not modelled as a catalog Assembly here: it is a device group, not a composed Microscope-style blueprint. (2-BM models its sample positioning as a SampleTower Assembly + Fixture; TomoWISE could earn the same once the manipulator firms and a scenario registers it.) So these pages carry the containment trees but no Assembly-to-Fixture composition diagram.
The model in one picture¶
The kinematic stacks, base to sample (containment, Asset.parent_id). The precise sub-order firms with the mechanical design (TDR manipulator tables 9.1 / 9.5); the trees below are the design-layout intent.
Microtomography (~45 m), mirroring the standard table -> tilt -> rotary -> sample-top order:
TomoWISE (Unit, Asset)
└── SampleTable (Component, Family Table; base, fixed at 45 m, Xt/Yt/beta)
├── LaminographyTilt (Device, TiltStage; tilts the rotation axis, alpha)
│ └── Rotary (Device, RotaryStage; wy, master rotation + trigger clock)
│ ├── SamplePositioning (Device, LinearStage; Xs/Zs, co-rotates)
│ └── SlipRing (Device, SlipRing; continuous-rotation feedthrough)
├── SampleSlit (Device, Slit; sample-side, beam-defining)
└── FastShutter (Device, Shutter; sample-side)
Nanotomography (~49 m), the granite carrying the KB optics and the coarse-to-fine manipulator:
TomoWISE (Unit, Asset)
└── NanoGranite (Component, Family Table; granite support, ~49 m)
├── KB (Device, Mirror; KB focusing pair)
└── NanoCoarseZ (Device, LinearStage; Zt, long-travel into the KB focus)
└── NanoCoarseY (Device, LinearStage; Yt, sample height)
└── NanoCoarseX (Device, LinearStage; Xt, centre-of-rotation)
└── NanoTilt (Device, TiltStage; Tilt X, axis-to-beam)
└── NanoRotary (Device, RotaryStage; Rot y, continuous rotation)
└── NanoSamplePositioning (Device, LinearStage; Xs/Zs, fine centring)
Microtomography endstation (~45 m)¶
The workhorse station: a fixed sample table about 45 m from the source carrying the rotation and positioning stack. The Rotary stage is also the trigger master clock (see Controls).
| Device | Family | Target model | Design spec (TDR) |
|---|---|---|---|
SampleTable |
Table |
(in-house) | fixed at 45 m, surface 390 mm below the beam; Xt +/-100 mm, Yt +50/-150 mm, beta tilt 1.2 deg |
Rotary |
RotaryStage |
RT100AX (STAGE-1) | 1200 rpm, 1 mdeg, TTL encoder 3600 pulses/rev; trigger master clock |
SamplePositioning |
LinearStage |
XY150B-12 (STAGE-2) | Xs/Zs centring, +/-6 mm per axis, 0.1 um |
LaminographyTilt |
TiltStage |
(target) | 25 deg tilt for laminography, distinct from tomography rotation |
SampleSlit |
Slit |
(target) | 50 x 5 mm aperture; JJ X-ray IB-C50-air reference design |
FastShutter |
Shutter |
(target) | sample-side fast shutter; Arinax Colibri (<5 ms) / Innospexion (<10 ms) references |
SlipRing |
SlipRing |
(target) | 30 to 40 channels for continuous-rotation acquisition up to 1000 rpm |
Optional modules the TDR anticipates (a horizontal-rotation loading rig for in-situ mechanics, a kHz tomography module) are not yet modelled; they join as confirmed.
Nanotomography endstation (~49 m)¶
The high-resolution station: the KB mirror pair focuses the undulator beam for 200-nm-class cone-beam imaging. The sample manipulator is a six-axis stack on a granite support, conceptually like the microtomography endstation but about ten times more precise; it reuses the same Families (no NanoPositioner of its own). The rotary is the critical axis: its Abbe error from wobble and eccentricity must not exceed 100 nm at 100 mm sample height. Each axis names a "(target)" model carried pending procurement (NANO-1).
| Device | Family | Target model | Design spec (TDR Table 9.5) |
|---|---|---|---|
KB |
Mirror |
(target) | KB pair, fixed-curvature graded-multilayer; focus 205 x 196 nm @ 30 keV, 196 x 80 nm @ 45 keV |
NanoGranite |
Table |
(target) | granite support housing the KB optics, the manipulator, and the detector stage |
NanoTilt |
TiltStage |
Huber 5202.80 | Tilt X, 2 deg, 5 mdeg; aligns the rotation axis to the beam |
NanoCoarseX |
LinearStage |
Huber 5101.20 | Xt coarse, 50 mm; CoR alignment + flat field |
NanoCoarseY |
LinearStage |
Huber 5103.A20-90 | Yt coarse, 50 to 100 mm; sample height |
NanoCoarseZ |
LinearStage |
Zaber X-LDQ-AE | Zt long-travel, 250 to 300 mm, <0.5 um; into the KB focus, then toward the detector |
NanoRotary |
RotaryStage |
RT100AS | Rot y, continuous, 1 mdeg, eccentricity <100 nm, TTL 3600/rev |
NanoSamplePositioning |
LinearStage |
XY150B-12 | Xs/Zs centring, +/-6 mm, 0.1 um |
The KB pair and the sample manipulator are the only nano-specific hardware; the shared beam delivery and detector serve both stations.
See Open questions for the model bindings still to confirm and Inventory for the Asset tree.