Source¶
The incident beam, produced, conditioned, and defined before the sample. A walk along the source-stage devices; the sample and detection stages are documented as their own composed-fixture pages, the controllers that drive these devices are on the Controls page, and the supplies they draw on are in Operations. Each device pairs its human name with the EPICS handle, its key specs, and whether it is field replaceable. new marks a device not yet modeled in CORA; confirm marks a value taken from the docs that staff have not yet verified.
Generated from the descriptor
This page is generated from the descriptor at deployments/tomowise/beamline.yaml. Edit the descriptor, not this page. For the CORA Asset model, settings, vendor catalog, drawings, and wiring, see Inventory.
| Property | Value |
|---|---|
| Facility | maxiv |
| Tier | Unit |
| Source | insertion-device (confirm: two switchable insertion devices: CPMU14 undulator + 3T3PW wiggler, selected per operation mode (TDR design phase)) |
| Z span | 0 to 52000 mm (confirm) |
Enclosures¶
| Enclosure | Role | Facility | Permit signal |
|---|---|---|---|
TomoWISE-optics |
optics-hutch | maxiv |
confirm: MAX IV PSS permit signal TBD (PSS-1) |
TomoWISE-experiment |
experiment-hutch | maxiv |
confirm: MAX IV PSS permit signal TBD (PSS-1) |
Insertion devices¶
Two switchable straight-section sources. The undulator drives high-flux micro- and nano-tomography; the wiggler drives wide-field large-FOV white-beam microtomography. The operation mode selects which is in use.
Enclosure: TomoWISE-optics.
Note
z is measured from the straight-section centre here (CPMU14 at -505 mm, 3T3PW at +1734 mm). All specifications are TDR design targets.
| Name | Family | PV | Key specs | Replaceable | Status |
|---|---|---|---|---|---|
CPMU14 |
InsertionDevice |
device type: cryogenic-permanent-magnet-undulator period: 14 mm min gap: 3.8 mm length: 2.0 m (cryostat 2.67 m) max k eff: 1.84 power at 500ma: 11.1 kW primary source for high-flux micro- and nano-tomography |
new confirm |
||
3T3PW |
InsertionDevice |
device type: three-pole-wiggler field: 3 T magnetic length: 412 mm min gap: 11 mm power at 500ma: 1.6 kW wide-field large-FOV white-beam microtomography |
new confirm |
Front end¶
Fixed and movable masks plus the heat absorber that shape the beam and absorb power before the optics hutch.
Enclosure: TomoWISE-optics.
Note
z from the CPMU14 source (TDR Table 7.1). Mask is now a shared catalog Family; the heat-absorber Family is not yet in the catalog (it renders as plain text pending the passive beam-path tier), bound loosely by design intent.
| Name | Family | PV | Key specs | Replaceable | Status |
|---|---|---|---|---|---|
FM1 |
Mask |
z mm: 14000 aperture: 1.100 mrad X/Y power absorbed: ~2 kW first fixed mask |
new confirm |
||
FM2 |
Mask |
z mm: 15000 aperture: 1.100 mrad X / 0.100 mrad Y power absorbed: up to 7 kW second fixed mask |
new confirm |
||
MSM |
Mask |
z mm: 16500 aperture: 1.1 x 0.1 / 0.1 x 0.1 mrad2 movable safety mask; <100 um lateral parasitic motion (design) |
new confirm |
||
HeatAbsorber |
HeatAbsorber |
z mm: 17000 power budget: 4 kW front-end heat absorber protecting the safety shutter; a second, near-identical heat absorber sits in the optics-hutch safety unit (SU) with SS1/SS2 (TDR Table 8.1), not separately modelled here |
new confirm |
Optics hutch¶
The conditioning optics that filter, monochromate, and define the beam delivered to the experiment hutch. Diagnostic modules DM1/DM2 carry the bremsstrahlung collimators, white-beam stops, slits, and beam-position monitors.
Enclosure: TomoWISE-optics.
Note
z from the straight-section centre (TDR Table 8.1). Filter / BPM / collimator / beam-stop Families are not all in the catalog yet and render as plain text.
| Name | Family | PV | Key specs | Replaceable | Status |
|---|---|---|---|---|---|
CVD |
Filter |
z mm: 24050 material: 0.35 mm CVD diamond transmission: 95% @ 20 keV, 50% @ 10 keV diamond filter |
new confirm |
||
WhiteBeamSlit |
Slit |
z mm: 24280 water-cooled white-beam slits (WBS) in DM1; DM1 also carries a separate beam-defining mask (BM, water-cooled Cu, 60 urad opening matching the MLM acceptance) and a bremsstrahlung collimator, not separately modelled here |
new confirm |
||
PFU |
Filter |
z mm: 25200 mechanism: two Si wedges, 3 deg wedge angle effective thickness: 0.2 to 25 mm power filters unit |
new confirm |
||
MLM |
Monochromator |
z mm: 25900 coating: W/SiC or W/B4C bilayers (TDR text); Table 8.3 lists W/Si; d=2.5 nm, 140 bilayers (OPT-1) energy range: 20 to 65 keV bandwidth: dE/E ~ 1.8% bragg range: 3.84 to 12.6 mrad max power: 100 W multilayer monochromator; bypassed in white-beam modes |
new confirm |
||
MonochromaticBeamSlit |
Slit |
z mm: 28280 monochromatic beam slits (DM2) |
new confirm |
||
MF |
Filter |
transmission: 1e-4 to 1 metal filter unit (third filter stage, Fe/Cu carriages, NanoMAX design); at the beginning of the experiment hutch |
new confirm |
Safety shutters¶
The safety unit (SU) gating the beam between the optics and experiment hutches.
Enclosure: TomoWISE-optics.
Note
z from the straight-section centre. PSS permit signals TBD (PSS-1).
| Name | Family | PV | Key specs | Replaceable | Status |
|---|---|---|---|---|---|
SS1 |
Shutter |
z mm: 31830 safety shutter (SU) |
new confirm |
||
SS2 |
Shutter |
z mm: 31830 safety shutter (SU) |
new confirm |